NICO
NICO

NICO

TriOS's new low-cost nitrate meter Based on the device platform concept of TriOS sensors like OPUS, LISA and VIPER, TriOS introduces NICO: a low-cost UV photometer for the determination of nitrate. The four detection channels enable a precise optical…  Read more

Measurement technology

light source: Xenon flash lamp

detector: 4 photo diodes + filter

Measurement principle Attenuation
Optical path 0.3 mm, 1 mm, 2 mm, 5 mm, 10 mm, 50 mm
Parameters NO3-N, NO3, NOx-N, NOx (calibrated with NO3 standard solution
Measurement range

1 mm path: 0.5...60 mg/L NO3-N

10 mm path: 0.05...6 mg/L NO3-N

Measurement accuracy

± (5 % + 0.1 mg/L NO3-N) with 10 mm path

± (5 % + 1 mg/L NO3-N) with 1 mm path

Turbidity compensation Yes
Data logger ~ 2 GB
T100 response time 20 s
Measurement interval ≥ 10 s
Housing material Stainless steel (1.4571/1.4404) or titanium (3.7035)
Dimensions (L x Ø)

~470 mm x 48 mm (with 10 mm path)

~ 18.5“ x 1.9“ (with 10 mm path)
Weight

stainless steel: ~ 3 kg

~ 6.6 lbs
titanium: ~ 2 kg ~ 4.4 lbs
Interface

digital:

Ethernet (TCP/IP)

RS-485 (Modbus RTU

Power consumption ≤ 7 W
Power supply 12...24 VDC (± 10 %)
Required supervision Typically ≤ 0.5 h/month
Calibration/maintenance interval 24 months
System compatibility Modbus RTU
Warranty

1 year (EU: 2 years)

US: 2 years

Installation

Max. pressure

with SubConn: 30 bar  

~ 435 psig

with fixed cable: 3 bar

~ 43.5 psig
in flow unit: 1 bar, 2...4 L/min ~ 14.5 psig at 0.5 to 1 gpm
Protection type

IP68

NEMA 6P

Sample temperature

+2...+40 °C

~ +36 °F to +104 °F

Ambient temperature

+2...+40 °C

~ +36 °F to +104 °F

Storage temperature

-20...+80 °C

~ -4 °F to +176 °F

Inflow velocity

0.1...10 m/s

~ 0.33 fps to 33 fps